CONTINUOUS NON INVASIVE BLOOD PRESSURE AMPLIFIER


The NIBP100D noninvasive blood pressure system provides a continuous, beat-to-beat, blNIBP100Dood pressure signal recorded from the fingers of a subject. The noninvasive blood pressure (NIBP) monitoring system uses a double finger cuff that is comfortable for the subject to wear and easy to place on the hand. The cuffs (included with system) come in three sizes to accommodate children through large adults. The system outputs a continuous blood pressure waveform that is similar to a direct arterial pressure waveform. The monitor displays values for systolic, diastolic, mean blood pressure, and heart rate.
  • Accurate noninvasive blood pressure values
  • Easy to use
  • Comfortable for subjects to wear
  • Real-time, continuous, noninvasive blood pressure NIBP100D cuff

The NIBP100D interfaces with an MP150 data acquisition system (or third-party data acquisition system), via a DA100C and TCI105 Interface Connector. The AcqKnowledge software displays the blood pressure signal, plus systolic, diastolic, mean blood pressure and heart rate. It will also provide a detailed beat-to-beat analysis of the blood pressure signal.

The NIBP100D is calibrated using a standard blood pressure cuff that is placed around the subject’s upper arm. The unit automatically takes a blood pressure measurement from the subject and uses the value for calibration purposes. During the calibration process the system locates the pulse at the finger and performs a partial occlusion. It will switch from one finger to the next during the course of the recording to relieve the pressure from the occluded finger. The interval between finger rotations is user-selectable and can be as long as 60 minutes. During the rotation, the system takes another calibration reading to ensure that values are accurate.

The system is very user friendly and the initial setup and calibration period takes less than three minutes—that time includes placing the cuff around the upper arm and the sensor on the fingers. Placing the finger sensor is as simple as sliding the subject’s fingers through the two cuffs.

The system employs a vascular unloading technique to measure blood pressure at the fingers. A refined version of the Penáz' principle is used to deliver a continuous noninvasive blood pressure signal. The method is based on concentrically interlocking control loops for accurate long-term readings of finger blood pressure. Learn more about the method:

Human non invasive blood pressure

 

 

Specifications:

 

The NIBP100D Noninvasive Blood Pressure Monitoring System is suitable for small children (~6 years) to large adults

For calibration, see NIBP100D Calibration. For complete specifications, see the NIBP100D User Manual under the "Resources" tab above.

Double Cuff Finger Sensor Size Diameter

  • L 24 - 28 mm dark red
  • M 18 - 24 mm Dark blue
  • S 10 - 18 mm Light blue

Blood Pressure Cuffs  Latex free

  • Child (12 – 19 cm) works starting at 20 kg (44 lbs), age 6 years
  • Small Adult (17 – 25 cm)
  • Adult (23 – 33 cm)
  • Large Adult (31 – 40 cm)

NIBP100D continuous noninvasive arterial pressure

  • Parameter classification
    • Sys, Dia, Mean [mmHg]
    • Pulse [bpm]
  • Measuring range
    • Sys: 40 - 250 mmHg (5.3 – 33.3 kPa)
    • Dia: 30 - 210 mmHg (4 - 28 kPa)
    • Mean: 35 - 230 mmHg (4 – 30.6 kPa)
    • Heart rate indication range 20-200 bpm
    • Accuracy ±5 mmHg (0,6 kPa)
  • Display resolution 1 mmHg (0.1 kPa)
  • Inflation pressure
    • Typ.: 120 mmHg (16 kPa)
    • Min.: 30 mmHg (4 kPa)
    • Max.: 300 ±10 mmHg (41.3 kPa ±1.3 kPa)
  • Excess pressure limit
    • 300 ±10 mmHg (40 kPa ±1.3 kPa)
    • Response time: < 3 sec.
    • Deflation time: < 15 sec
    • Protection against electric shock: Type BF

Interface

  • To DA100C via TCI105 (sold separately)

Output

  • Sensor bridge voltage: 2 - 10V (external monitor)
  • Sensitivity: 5 µV/V/mmHg
  • BP Wave Out: CNAP™ transducer cable 0.3m; connector RJ11 6P4C (e.g. Abbott IBP catheter)
  • Maximum delay of analog out signal: 80 msec (means the BP waveform may be delayed with respect to other waveforms acquired by the MP150 unit, such as ECG, by max 80 msec)

NIBP100D Monitor

  • Dimensions 280 x 270 x 250 mm (11 x 10.6 x 9.8 in.)
  • Weight 7.5 Kg (16.6 lbs) including components and accessories necessary for operability of device
  • Battery Sealed lead gel, operating time = 2 hours (fully charged battery, normal conditions)

Electrical properties

  • Nominal voltage: 18 VDC ±10%
  • Nominal current: 3A
  • Operability: No time-limit if powered by external mains adapter, at least 2h if on battery-operation (fully charged battery)

External mains adapter

  • Nominal voltage: 100 – 240 VAC
  • Power frequency: ~50/60 Hz
  • Power output: 18 V, 3.3 A
  • Safety class: Class II with functional earth
  • Earth leakage current: < 500 µA

Note—Electric and magnetic fields may interfere with the functional reliability of the device, so avoid using the CNAP™ Monitor 500 close to devices emitting powerful electromagnetic fields, e.g. x-ray equipment, diathermy applications or magnetic resonance tomographs.

 

Part Numbers:

Item Description
NIBP100D Continuous Non Invasive Blood Pressure Amplifier
DA100C Amplifier Interface for NIBP100D
TCI105 Transducer Connector Interface

 

Note:BIOPAC Systems, Inc., instruments, components, and accessories are designed for educational- and research-oriented life science applications and investigations. BIOPAC Systems, Inc., does not condone the use of its instruments for clinical medical applications. Instruments, components, and accessories provided by BIOPAC Systems, Inc., are not intended for the diagnosis, cure, mitigation, treatment, or prevention of disease.